Test Code:
090-20-2020-01

Order Name:
Sweat Conductivity Test

 
Useful For:
Screening for Cystic fibrosis
 
Methodology:
Conductivity
 
 
 
Test Code:
090-20-2020-01

Order Name:
Sweat Conductivity Test

 
Collection Specimen Or Container:
Sweat, Special container
 
Specimen Testing Type:
Sweat, minimum 15 µL
 
Rejection Criteria:
  1. Specimen volume less than 15 µL
  2. The patient who use the open delivery oxegen system
  3. The Serious respiratory diseases patient (TB , MERS)
  4. The Skin infection patient 5. HIV infection patient
 
Specimen Stabillity:
Sweat (2oC to 25oC and keep in microcentrifuge tube size 0.2 mL, close the lid tightly) - within 72 hours after collection
 
 
 
Test Code:
090-20-2020-01

Order Name:
Sweat Conductivity Test

 
Method detail:
Conductivity
 
Schedule:
Make an appointment with the laboratory.
 
Turnaround Time:
Collected specimen to report within 1.5 hours (90 mins)
 
Performing Location:
Chemistry, Laboratory Department Tel. 13224
 
Specimen Retention Time:
1 month
 
 
 
Test Code:
090-20-2020-01

Order Name:
Sweat Conductivity Test

 
 
Clinical Information:
Cyctic fibrosis (CF) it the common life-threatening genetic disease within the white population. It id an autosomal recessive disorder characterized by viscous secretions that affect the exocrine glands, primarily in the lung and pancreas. Patient with CF have an increased concentration of sodium, chloride and potassium in thier sweat. The criteria for diagnosis of CF include the presence of one or more characteristic phenotypic features, or a history of CF in a sibling, or a positive new born screening test result; and an increased sewat chloride concentration by pilocarpine iontophoresis on two or more occasions, or identification of two CF-causing mutations or demonstration of abnormal nasal epithelial ion transport.
 
Reference Value:
0 -50 mmol/L
 
Clinical Reference:
  1. SWEAT-CHEKTM Sweat Conductivity Analyzer Model 3120 Instrument Service / Manual, Wescor, Inc. 2005.
  2. Sweat Testing: Sample Collection and Quantitative Analysis; Approved Guideline – Second Edition.
  3. NCCLS document C34-A2 (ISBN 1-56238-407-4). NCCLS, Pennsylvania USA, 2000.